1.5 Testing Sensors
1. Before daily operations and deployments, verify the instrument clicking
Survey. Select all installed sensors and click Start Survey. Click
Open Plot to view the data .
2. Wave your hand in front of the OBS sensor; the turbidity signal on the top
plot of FIGURE 1-7 will fluctuate and data will scroll.
3. Blow on the temperature sensor to observe an increase in temperature (red
trace on the middle plot of FIGURE 1-7).
4. Dip the pressure sensor in water and a small elevation in the pressure
signal will occur (bottom plot of FIGURE 1-7).
5. Dip the sensor in salty water and conductivity will increase (blue trace on
middle plot).
6. Click
OBS-3A Info to view time, serial numbers, depth corrections,
and software versions (FIGURE 1-8).