PWC100 Present Weather Sensor Calibrator
Step 4: Position 2: Left
Move the calibrator into the left hand position by unscrewing the calibrator
mounting screws slightly and sliding the calibrator fully to the left. Then
retighten the screws once in position.
Visually check that the calibrator is located correctly and securely.
LED L should now be flashing slowly and all other LEDs should be off.
Once you are ready, press START.
The test is now automated and the disk should start rotating. LED L will
stay illuminated until this stage of the test is complete.
Wait for the calibration disk to stop rotating. This test will take
approximately two and a half minutes.
Upon successful completion of the test, LED R will be flashing slowly and
LED L will be turned off.
7