Campbell-scientific CS110 Electric Field Meter Manuel d'utilisateur Page 77

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 88
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 76
Appendix G. CS110 2 Meter CM10 Tripod Site
Site Correction for Tripod CS110 Only Site - October 2, 2005
1 minute average of 1 second data
Results indicate Csite = 0.105.
y = 0.1051x - 35.664
R
2
= 0.9996
-10000
-8000
-6000
-4000
-2000
0
2000
4000
6000
8000
-80000 -60000 -40000 -20000 0 20000 40000 60000 80000
Uncorrected (Csite = 1) Electric Field (volt/meter) for Tripod CS110 Only site
SN: 1023 Mparallel_Plate = 81.77 volt/meter*millivolt
Electric Field (volt/meter for Upward Facing CS110
SN:1022 Mparallel_plate = 87.6 volt/meter*millivolt
Tripod CS110 Only Uncorrected E_Field 1_Min_Avg volts/m
Linear (Tripod CS110 Only Uncorrected E_Field 1_Min_Avg
volts/m)
FIGURE G-4. Determination of Csite
G-5