Campbell-scientific CS547A-L Conductivity/Temperature Probe and A547A Manuel d'utilisateur

Naviguer en ligne ou télécharger Manuel d'utilisateur pour Instruments de mesure Campbell-scientific CS547A-L Conductivity/Temperature Probe and A547A . Campbell Scientific CS547A-L Conductivity/Temperature Probe and A547A Interface User Manual [es] Manuel d'utilisatio

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 30
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 0
CS547A Conductivity and
Temperature Probe and
A547 Interface
Revision: 10/11
Copyright © 1994-2011
Campbell Scientific, Inc.
Vue de la page 0
1 2 3 4 5 6 ... 29 30

Résumé du contenu

Page 1 - A547 Interface

CS547A Conductivity andTemperature Probe andA547 InterfaceRevision: 10/11Copyright © 1994-2011Campbell Scientific, Inc.

Page 2

CS547A Conductivity and Temperature Probe and A547 Interface Datalogger AGSE TEMPEX TEMPDATALOGGER SENSOREX TEMPTEMPCONDEX CONDEX CONDHI CONDLO CONDSH

Page 3 - Warranty

CS547A Conductivity and Temperature Probe and A547 Interface 2. Correction of ionization errors in EC measurements Ionization caused by the excitati

Page 4 - Assistance

CS547A Conductivity and Temperature Probe and A547 Interface '\\\\\\\\\\\\\\\\\\\\\\\\\\\ PROGRAM //////////////////////////// BeginProg '

Page 5 - Table of Contents

CS547A Conductivity and Temperature Probe and A547 Interface 'correct errors in the EC measurement due to temperature Therm107 (TempDeg_C,1,3

Page 6

CS547A Conductivity and Temperature Probe and A547 Interface 6: BR Transform Rf[X/(1-X)] (P59) 1: 1 Reps 2: 1 Loc [ Rs ] 3

Page 7 - Probe and A547 Interface

CS547A Conductivity and Temperature Probe and A547 Interface ; ;Subtract resistance errors (Rp) caused by the blocking capacitors ;(0.005Kohm) and the

Page 8 - 2. Specifications

CS547A Conductivity and Temperature Probe and A547 Interface 24: IF (X<=>F) (P89) 1: 3 X Loc [ Ct ] 2: 4 < 3: .474 F 4

Page 9 - 4. Wiring

CS547A Conductivity and Temperature Probe and A547 Interface 33: Z=X*F (P37) 1: 6 X Loc [ A ] 2: nnn F Enter TC (%/°C) to correct

Page 10 - 5. Programming

CS547A Conductivity and Temperature Probe and A547 Interface 6. Calibration 6.1 Conversion Factors 1 S (Siemens) = 1 mho = 1/ohm Although mS·cm-1 an

Page 11 - 5.2.1 CRBasic

CS547A Conductivity and Temperature Probe and A547 Interface The calibration solution temperature must be between 1°C and 35°C; a polynomial is used t

Page 13 - 5.2.2 Edlog

CS547A Conductivity and Temperature Probe and A547 Interface 2: BR Transform Rf[X/(1-X)] (P59) 1: 1 Rep 2: 1 Loc [Rs ] 3: 1 Mul

Page 14

CS547A Conductivity and Temperature Probe and A547 Interface 11: Polynomial (P55) 1: 1 Rep 2: 3 X Loc [T25_01] 3: 4 F(X) Loc [f_of_T

Page 15

CS547A Conductivity and Temperature Probe and A547 Interface derived by measuring the standard solutions described in Section 2.2 with values of 0.023

Page 16

CS547A Conductivity and Temperature Probe and A547 Interface 8.2 Temperature Measurement Error The overall probe accuracy is a combination of the the

Page 17

CS547A Conductivity and Temperature Probe and A547 Interface 9. Deriving a Temperature Compensation Coefficient 1. Place the CS547A in a sample of t

Page 18 - 6. Calibration

CS547A Conductivity and Temperature Probe and A547 Interface TABLE 10-1. Temperature , Resistance, and Datalogger Output 0.00 351017 -0.06 2.00 315

Page 19

CS547A Conductivity and Temperature Probe and A547 Interface 11. Electrically Noisy Environments AC power lines can be the source of electrical noise

Page 20

CS547A Conductivity and Temperature Probe and A547 Interface Example 12-1. CR1000 measurement instruction with 20 mSec (20000 uSec) delay: Therm107(

Page 21 - 8. Analysis of Errors

CS547A Conductivity and Temperature Probe and A547 Interface 22 FIGURE 13-2. A547 Interface circuit diagram EX TEMP EX COND COND TEMP SHIELD EX

Page 23

Warranty “PRODUCTS MANUFACTURED BY CAMPBELL SCIENTIFIC, INC. are warranted by Campbell Scientific, Inc. (“Campbell”) to be free from defects in materi

Page 24

Campbell Scientific Companies Campbell Scientific, Inc. (CSI) 815 West 1800 North Logan, Utah 84321 UNITED STATES www.campbellsci.com • info@campbel

Page 25

Assistance Products may not be returned without prior authorization. The following contact information is for US and international customers residing

Page 26

CS547A Probe and A547 Interface Table of Contents PDF viewers: These page numbers refer to the printed version of this document. Use the PDF reader

Page 27 - Orange (Cond)

CS547A Probe and A547 Interface Table of Contents ii 10. Therm107 / P11 Instruction Details ...18 11. Electrically Noisy Enviro

Page 28

CS547A Conductivity and Temperature Probe and A547 Interface 1. Overview The CS547A conductivity and temperature probe, and A547 interface are design

Page 29

CS547A Conductivity and Temperature Probe and A547 Interface FIGURE 1-1. A547 Interface and CS547A Conductivity and Temperature Probe 2

Page 30

CS547A Conductivity and Temperature Probe and A547 Interface 2.3 Temperature Sensor Thermistor Betatherm 100K6A1. Range 0° to 50°C. Accuracy Error ±0

Commentaires sur ces manuels