Campbell-scientific CS547 Conductivity and Temperature Probe and A547 Manuel d'utilisateur Page 17

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 26
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 16
CS547 Conductivity and Temperature Probe and A547 Interface
13
Correction of Ionization Errors: Figures 8.1-1 and 8.1-2 show the amount of
correction applied by the example program to compensate for ionization effects on
the measurements. Also shown is an ideal correction. Factors were derived by
measuring the standard solutions described in Section 2.2 with values of 0.0234,
0.07, 0.4471, 07, 1.413, 2.070, 3.920, and 7.0 mS cm
-1
.
FIGURE 8.1-1. Plot of Ideal and Actual Correction between
0 and 0.44 mS cm
-1
FIGURE 8.1-2. Plot of Ideal and Actual Correction between
0.44 and 7.0 mS cm