CS650 and CS655 Water Content Reflectometers
6.4.2.1 Settings Editor Tab
The Settings Editor tab shows settings stored in the CS650 firmware. Settings
that may be modified include User Name, SDI-12 Address, and RS-232 Baud
Rate. Attempts to change any of the other settings will result in a “Commit
failed. Unrecognized error condition” error message. DevConfig polls the
CS650 every two seconds while connected and the results are displayed in the
Real-Time Measurements field. This is useful for verifying probe
performance.
After any changes to CS650 settings, select Apply to write the changes to the
CS650 firmware. A configuration summary is then shown. The summary may
be printed or saved electronically for future reference.