HMP35C TEMPERATURE AND RH PROBE
6
Example 2. Sample CR10(X) Program measuring HMP35C Temperature
using DC Half Bridge with Delay
;Measure the HMP35C thermistor.
;
01: Excite-Delay (SE) (P4)
1: 1 Reps
2: 2 7.5 mV Slow Range
;CR500 (7.5 mV); CR23X (10 mV); 21X, CR7 (15 mV)
3: 3 SE Channel
;Orange wire (SE 3), White wire (AG)
4: 1 Excite all reps w/Ex 1
;Black wire (EX 1)
5: 2 Delay (units 0.01 sec)
6: 2000 mV Excitation
;CR500, CR23X (2000 mV); 21X, CR7 (4000 mV)
7: 1 Loc [ T_C ]
8: 0.4 Mult
;CR500, CR23X (0.4); 21X, CR7 (0.2)
9: 0 Offset
;Compute the temperature in degrees Celsius.
;
02: Polynomial (P55)
1: 1 Reps
2: 1 X Loc [ T_C ]
3: 1 F(X) Loc [ T_C ]
4: -53.46 C0
5: 90.807 C1
6: -83.257 C2
7: 52.283 C3
8: -16.723 C4
9: 2.211 C5
;Apply the temperature multiplier.
;
03: Z=X*F (P37)
1: 1 X Loc [ T_C ]
2: 1 F
;See Table 4 for alternate multipliers
3: 1 Z Loc [ T_C ]
;Apply the temperature offset.
;
04: Z=X+F (P34)
1: 1 X Loc [ T_C ]
2: 0 F
;See Table 4 for alternate offsets
3: 1 Z Loc [ T_C ]
;Turn the HMP35C on.
;
05: Do (P86)
1: 41 Set Port 1 High
;Yellow wire (C1)
;Pause 150 milliseconds, before making the measurement, so the
;probe can stabilize on the true relative humidity.
;
06: Excitation with Delay (P22)
1: 1 Ex Channel
2: 0 Delay W/Ex (units = 0.01 sec)
3: 15 Delay After Ex (units = 0.01 sec)
4: 0 mV Excitation