NR-LITE2 Net Radiometer
3. Select NR-LITE2 Net Radiometer (no wind speed correction), and
select the right arrow (in center of screen) to add it to the list of sensors to
be measured, and then select Next.
4. Define the name of the public variable and enter the calibration factor.
The public variable defaults to NR_Wm2. The calibration factor is unique
to each sensor. This value is provided on the certification of calibration
that is shipped with your sensor. After entering the information, click on
OK, and then select Next.
6