33: Z=X*F (P37)
1: 6 X Loc [ A ]
2: nnn F Enter TC (%/
°
C) to correct cond. reading.
3: 8Z Loc [ TC_Proces ]
34: Z=X+F (P34)
1: 8 X Loc [ TC_Proces ]
2: 100 F
3: 8 Z Loc [ TC_Proces ]
35: Z=X/Y (P38)
1: 7 X Loc [ Ct100 ]
2: 8 Y Loc [ TC_Proces ]
3: 9 Z Loc [ C25mScm_l ] EC corrected for temperature.
;Output processing, convention states that the temperature be reported
;with the EC measurement.
;
36: Do (P86)
1: 10 Set Output Flag High (Flag 0)
37: Sample (P70)
1: 1 Reps
2: 3 Loc [ Ct ]
38: Sample (P70)
1: 1 Reps
2: 4 Loc [ Temp_degC ]
39: Sample (P70)
1: 1 Reps
2: 9 Loc [ C25mScm_l ]
*Table 2 Program
02: 0.0 Execution Interval (seconds)
*Table 3 Subroutines
End Program