CSAT3 Three Dimensional Sonic Anemometer
FIGURE 11-4. Recommended settings for the wind vertical axis
FIGURE 11-5. Recommended settings for the wind vertical axis labels
Collect and graph 1 minute of wind data from the CSAT3 while it is in the zero
wind environment (FIGURE 11-6). The wind offset should be less than ± 4 cm
s
–1
(0.04 m s
–1
) for u
x
and u
y
, and less than ± 2 cm s
–1
(0.02 m s
–1
) for u
z
. If the
CSAT3 wind offset is greater than these specifications, contact Campbell
Scientific, Inc. for an RMA number to recalibrate the CSAT3.