Campbell-scientific CS470/CS471 Compact Bubbler System Manuel d'utilisateur Page 28

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 46
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 27
CS470/CS471 Compact Bubbler System
Any error states arising can be displayed as follows:
Press pump membrane button briefly (< 1 sec). The LED lights
once for a longer period as confirmation, then pauses. 1
st
error
state arising, then pause. 2
nd
error state arising, then pause. LED
lights continue until all error states arising have flashed. The
CS470/CS471 repeats all error states arising for approximately
two minutes.
The most common causes for erroneous data include:
poor sensor connections to the datalogger
damaged cables
damaged transducers
Problem:
Unit will not respond when attempting serial communications.
Suggestion:
Check the power and signal lines to ensure proper connection to the datalogger.
Check the datalogger program to ensure that the same port the SDI-12 data line
is connected to is specified in the measurement instruction.