Section 7. Installation
164
'AVW200(AVWRC,Com1,0,200,VW(1,1),1,1,1,1000,4000,1,_60Hz,1,0) '<<actual measurement
'instruction (commented out)
'Digits_Measured=(VW(1,1)^2)/1000 '<<actual processing expression (commented out)
Digits_Measured = 8746
'Temp_Measured=1/(1.4051E-3 + 2.369E-4 * LN(VW(1,6))+1.019E-7 * LN(VW(1,6)) ^3)-273.15
Temp_Measured = 21.4
'VoltSE(BP_Measured,1,mV2500,5,1,0,_60Hz,0.2,600) '<<actual measurement instruction
'(commented out)
BP_Measured = 991
FieldCal(4,Equation_Parameters(),3,0,Offset(),CalMode,0,1,1)
Pressure = (GageFactor * (Digits_Measured - Digits_Offset) + Temp_K * _
(Temp_Measured - Temp_Offset) - (BP_Measured - BP_Offset) * 0.014503)
NextScan
EndProg
7.8.1.5.4 Two-Point Slope and Offset (Option 2)
Case: A meter measures the volume of water flowing through a pipe. Multiplier
and offset are known to drift, so a two-point calibration is required periodically at
known flow rates. The following procedure adjusts multiplier and offset to correct
for meter drift as shown in the calibration report below. Note that the flow meter
outputs milliVolts inversely proportional to flow.
Table 26. Calibration Report for Flow Meter
Parameter
Parameter at
Deployment
Parameter at 7 Day
Service
SignalmV output @ 30
L/s
300 mV 285 mV
SignalmV output @ 10
L/s
550 mV 522 mV
Multiplier
-0.0799 L/s/mV -0.0841 L/s/mV
Offset
53.90 L 53.92 L
1. Send the program in CRBasic example FieldCal Multiplier and Offset
Demonstration Program
(p. 165) to the CR3000.
2. To simulate the flow sensor, place a jumper wire between channels VX1 and
SE6 (3L).
3. Simulate deployment-calibration conditions (output @ 30 l/s = 300 mV, output
@ 10 l/s = 550 mV) in two stages.
a. Set variable SignalmV to 300. Set variable KnownFlow to 30.0.
b. Start the deployment calibration by setting variable CalMode = 1.
c. When CalMode increments to 3, set variable SignalmV to 550. Set
variable KnownFlow to 10.
d. Resume the deployment calibration by setting variable CalMode = 4
Vue de la page 163
1 2 ... 159 160 161 162 163 164 165 166 167 168 169 ... 589 590

Commentaires sur ces manuels

Pas de commentaire

Philips HTR5204 manuals

Owner’s manuals and user’s guides for CD players Philips HTR5204.
We providing pdf manuals Philips HTR5204 for download free by document types:






More products and manuals for CD players Philips

Models Document Type
DVDQ40/022 User Manual    Philips DVDQ40/022 User's Manual, 26 pages
AZ382/12 User Manual   Philips AZ382/12 User's Manual, 2 pages
MZ1000 User Manual   Philips MZ1000 User's Manual, 25 pages
EXP4318 User Manual      Philips EXP4318 User's Manual, 82 pages
AZT3201 User Manual      Philips AZT3201 User's Manual, 10 pages
ACT 7585 User Manual      Philips ACT 7585 User's Manual, 65 pages
AZ3012/01 User Manual    Philips AZ3012/01 User's Manual, 8 pages
eXpanium EXP212 User Manual      Philips eXpanium EXP212 User's Manual, 2 pages
AZ 1565/00 User Manual       Philips AZ 1565/00 User's Manual, 63 pages
AZT3200 User Manual     Philips AZT3202/17 User manual, 14 pages
EXP2550/58 User Manual      Philips EXP2550/58 User manual, 2 pages
EXP3360 User Manual   Philips EXP3360 User's Manual, 12 pages
AZ7681 User Manual       Philips AZ7681 User's Manual, 132 pages
EXP220/01 User Manual   Philips EXP220/01 User manual, 5 pages
PORTABLE SPORT AUDIO PSA[CD User Manual      Philips PORTABLE SPORT AUDIO PSA[CD User's Manual, 14 pages
AX5311/17 User Manual   Philips AX5311/17 User manual, 13 pages
AZ7182 User Manual   Philips AZ7182 User's Manual, 2 pages
AZ2000/01 User Manual   Philips AZ2000/01 User's Manual, 11 pages
AX5301/00C User Manual   Philips AX5301/00C User's Manual, 2 pages
AZ 1145 User Manual      Philips AZ 1145 User's Manual, 53 pages