Section 8. Operation
303
'Main Program
BeginProg
R2 = 1000 'Resistance of R2
R3 = 1000 'Resistance of R3
R4 = 1000 'Resistance of R4
Scan(500,mSec,1,0)
'Full Bridge Measurement:
BrFull(X,1,mV2000,1,1,1,2500,True,True,0,_60Hz,1.0,0.0)
X1 = ((-1 * X) / 1000) + (R3 / (R3 + R4))
R1 = (R2 * (1 - X1)) / X1
NextScan
EndProg
8.1.3.1 ac Excitation
Some resistive sensors require ac excitation. These include electrolytic tilt
sensors, soil moisture blocks, water conductivity sensors, and wetness sensing
grids. The use of dc excitation with these sensors can result in polarization, which
will cause erroneous measurement, shift calibration, or lead to rapid sensor decay.
Other sensors, e.g., LVDTs (linear variable differential transformers), require an
ac excitation because they rely on inductive coupling to provide a signal. dc
excitation will provide no output.
CR3000 bridge measurements can reverse excitation polarity to provide ac
excitation and avoid ion polarization.
Note Sensors requiring ac excitation require techniques to minimize or eliminate
ground loops. See Ground Looping in Ionic Measurements (p. 96).
8.1.3.2 Accuracy of Ratiometric-Resistance Measurements
The ratiometric-accuracy specification for resistance measurements is:
±(0.02%*V1+Offset),to40°C,
±(0.025%*V1+Offset),‐25°to50°C,
±(0.03%*V1+Offset),‐40°to85°C,
whereV1isthevoltagemeasurementandOffsetisequaltooneofthe
following,wheretheBasicResolutionistheresolutionofasingleA/
D(p.
449)
conversion.Notethatexcitationreversalreducesoffsetsbyafactor
oftwo:
Offset = 1.5 x Basic Resolution + 1.0 µV if the measurement is made on a
differential input channel with input reversal
Offset = 3 x Basic Resolution + 2.0 µV if the measurement is made on a
differential input channel without input reversal