Section 8. Operation
315
Error Calculations with Input Reversal = False
µVerror=gainterm+offsetterm
=(44500µV*0.07%)+(3*6.67µV+2.0µV)
=31.2µV+22.0µV
=1.52µV(=1.52°C)
8.1.4.1.4 Ground Looping Error
When the thermocouple measurement junction is in electrical contact with the
object being measured (or has the possibility of making contact), a differential
measurement should be made to avoid ground looping.
8.1.4.1.5 Noise Error
The typical input noise on the ±20-mV range for a differential measurement with
16.67 ms integration and input reversal is 0.19 V RMS. On a type-T
thermocouple (approximately 40 V/°C), this is 0.005°C.
Note This is an RMS value; some individual readings will vary by greater than
this.
8.1.4.1.6 Thermocouple Polynomial Error
NIST Monograph 175 gives high-order polynomials for computing the output
voltage of a given thermocouple type over a broad range of temperatures. To
speed processing and accommodate the CR3000 math and storage capabilities,
four separate 6th-order polynomials are used to convert from volts to temperature
over the range covered by each thermocouple type. The table Limits of Error on
CR3000 Thermocouple Polynomials
(p. 315) gives error limits for the thermocouple
polynomials.
Table 69. Limits of Error on CR3000 Thermocouple Polynomials
TC
Type
Range °C
Limits of Error °C
Relative to NIST
Standards
T
-270 to 400
-270 to -200 +18 @ -270
-200 to -100 ±0.08
-100 to 100 ±0.001
100 to 400 ±0.015
J
-150 to 760
±0.008
-100 to 300 ±0.002
E
-240 to 1000
-240 to -130 ±0.4